VersaMed iVent201 Portable Ventilator Service Manual β PDF Download
$29.95
Official factory service manual for the VersaMed iVent201 mechanical ventilator (Hardware v1.4, Rev. 4, 2005).
Includes calibration, testing, maintenance, pneumatic and electronic schematics, troubleshooting, and full service procedures.
Description
VersaMed iVent201 Portable Ventilator Service Manual β PDF Download
DescriptionΒ
Overview
The VersaMed iVent201 Service Manual (Part No. SM-01-04 Rev.4, August 2005) provides the complete technical documentation required to service, calibrate, and verify the iVent201 portable ventilator. This manual is designed for authorized VersaMed technicians and biomedical engineers responsible for preventive maintenance and repair. It complements the Operatorβs Manual (OM-01-04) and contains all electrical, pneumatic, and software servicing procedures
69164909-Service-Manual
IMAGES:

Contents Summary
Section 1 β Introduction & Safety: Intended use, safety standards (IEC 60601-1/2-12, ISO 10651-3), ESD handling, hazard notices, and replacement part guidelines.
Section 2 β System Specifications:
Ventilation modes: A/C Volume, A/C Pressure, SIMV Volume, SIMV Pressure, CPAP, PSV, Adaptive Bi-Level.
Range highlights: VT 100β2000 mL Β±10%, FiOβ 21β100%, PEEP 0β20 cmHβO, Power 100β240 V AC or 12β15 V DC.
Dimensions 33 Γ 24 Γ 26 cm; weight 10 kg (with battery).
69164909-Service-Manual
Section 3 β Installation & Setup: External AC/DC supply, internal battery charging, oxygen connection (40β60 psi HP supply / low-pressure adapter), patient circuit assembly, filters, and control panel functions.
Section 4 β Theory of Operation:
Pneumatic Unit β blower assembly (22,000 RPM turbine), Oβ blending system (demand valve, proportioning valve with stepper motor control), solenoid valves, filters, and mufflers.
Electronic Module β CPU board (AMD 5Γ86), main board sensor interfaces, zero/purge and switching boards, power supply, and SpOβ interface.
Control Software β BIOS, watchdog timer, PC104 bus, RS-232 communication, and VGA output
69164909-Service-Manual
Section 5 β Maintenance & Calibration: Routine cleaning, PM intervals (500 h, 1500 h, annual), and sensor calibration procedures (flow, pressure, PEEP, Oβ system).
Section 6 β Verification Tests: Operational Verification (OVT), Ventilator Verification (VVT), Functional Verification Tests for Oβ linearity, alarms, and battery performance.
Section 7 β Service Procedures: Software upgrade, option package update, download of technical logs, and complete disassembly instructions for modules (power pack, LCD, interface board, pneumatic unit).
Section 8 β Troubleshooting: Detailed fault tables for power, pressure, valve, sensor, and Oβ control issues with root-cause isolation and waveform examples
69164909-Service-Manual
Technical Highlights
| Function | Specification |
|---|---|
| Ventilation Modes | A/C (V/P Ctrl), SIMV (V/P Ctrl), CPAP, PSV, Adaptive Bi-Level |
| Power Input | 100β240 V AC (1.6 A) / 12β15 V DC (8.5 A) |
| Battery Type | 12 V 7.2β9 Ah Sealed Lead-Acid, 2 hr runtime |
| Oβ Input Pressure | 40β60 psi (2.8β4.1 bar) |
| Operating Temp. | β15 to +50 Β°C (5β122 Β°F) / IP54 protection |
| Monitoring | FiOβ, PIP, PEEP, MV, Rate, I:E, Flow, Leak, Alarms (Apnea, Disconnect, High PEEP) |
| CPU Module | AMD 5Γ86-133 MHz PC104 Architecture + 32 MB SDRAM |
| Display | 8.4 inch LCD with rotary encoder and keypad controls |
π File Details
| Parameter | Information |
|---|---|
| File Type | PDF (Digital Download) |
| File Size | β 3.74MB |
| Page Count | β 173 pages |
| Language | English |
| Edition | Part No. SM-01-04 Rev. 4 β August 2005 |
| Searchable Text | Yes (OCR enabled) |
| Compatibility | Windows / macOS / iOS / Android |
| Publisher | VersaMed Medical Systems Inc. (USA) |
What Our Customers Say
π Related Products