★ TRUSTED BY 321,897+ PROFESSIONALS WORLDWIDE
🏢
Trusted Business
Verified & Licensed
🛡️
Virus Free Files
100% Safe Downloads
🔒
Secure Payment
SSL Protected
Instant Delivery
Available Immediately
Sale!

Biocode Hycel Diana 5 Evolution Xenia Technical Manual 576CA70E April 2006

Original price was: $34.95.Current price is: $19.95.

Technical manual for the Biocode Hycel Diana 5 Evolution and Xenia hematology analyzers covering installation, counting fluidics, five-part WBC differentiation, cytometer service, troubleshooting, and board wiring. Document reference 576CA70E, modified 19 April 2006.

Instant PDF Download
Available immediately
💾
Save to Your Device
Download & keep forever
🛡️
Antivirus Scanned
100% virus-free
🌍
Trusted Worldwide
175,000+ customers

Description

Biocode Hycel Diana 5 Evolution Xenia Technical Manual 576CA70E April 2006

This Biocode Hycel Diana 5 Evolution Xenia Technical Manual 576CA70E April 2006 is written for service engineers and laboratory technicians who install, calibrate, and keep the Diana Évo / Xénia hematology analyzer running. The instrument measures CBC parameters and a five-part leukocyte differential by combining impedance counting with hydrodynamic focusing and wide-angle light scatter in a dedicated cytometer. The text follows the real work sequence: site preparation and reagent hookup, compiled software sequences, every fluidic counting cycle, quality-control logic, scattergram interpretation, maintenance-program tests, cytometer disassembly, alarm-by-alarm troubleshooting, scheduled part exchanges, and printed wiring diagrams for the mother, sampling, sequence, power, counting, and hemoglobin boards.

File Details

  • Manual type: Technical manual
  • Brand / models: Biocode Hycel Diana 5 Evolution (Diana Évo) and Xénia
  • Language: English (installer screens and some labels also appear in French)
  • Page count: 414
  • File format: PDF
  • Document reference: 576CA70E
  • Revision note: Technical manual modified 19/4/06 (cover dated April 2006)

Instrument, Reagents, and Site Requirements

The analyzer plus compressor is specified for 115 V ± 10 % (105–125 V) or 230 V ± 10 % (205–245 V) at a maximum of 1000 VA. Ground quality is treated as mandatory: the ground line must not be interrupted between mains and the instrument plug, and voltage between neutral and ground must stay within 0.5 V. The optional autosampler is listed separately at 115/230 V and 150 VA. Ambient limits are 18 to 30 °C (64 to 86 °F) without condensation. Planned bench space is 1.50 m wide by 0.80 m deep, with a rear opening for fluidic and electrical lines. Published dimensions and weights are 54 × 56 × 56 cm and 40 kg for the analytical module and 42 × 17 × 60 cm and 18 kg for the compressor.

Reagent installation is color-coded. Detergent HEMACORE uses the blue Ellitec sampling assembly and level detector. Diluent HEMATON-5 uses the white connectors (three white fittings plus level detector). Lysing agent HEMALYSE-5 is a small tube with a green Luer fitting into a pierced flask cork. Yellow check valves on the analytical module and compressor both route to the waste container. Waste uses the orange Ellitec fittings and a cap with level detection. Rigid blue pipes carry vacuum and pressure between compressor and module; the black sleeve marks the pressure line (black to black-background fitting). Two electrical cables handle compressor power and command signals.

Host-computer guidance in this Biocode Hycel Diana 5 Evolution Xenia Technical Manual 576CA70E April 2006 lists a minimum Pentium II 200 MHz PC with two serial ports, 4 GB disk, 256 MB RAM, 800 × 600 display, diskette and CD-ROM drives, Windows 2000 or Windows XP Pro, and Internet Explorer 5. Recommended hardware is a 350 MHz Pentium II, 512 MB RAM, and a 17-inch 1024 × 768 display. The installer can choose unit systems (US, SI, SI2, Int’l 1, Int’l 2) that change how WBC, RBC, platelets, Hb, Hct, MCH, and MCHC are printed. A Compile utility turns .zzz sources plus the instrument diskette file param-seq.h60 into working sequence files for one analyzer. Software habilitation is split into operator, supervisor, and service accounts defined at first startup.

How a Count Cycle Is Built

Chapter 2 treats the analyzer as a multitasking fluidic machine. After cap piercing, blood is drawn in two segments separated by a bubble. The first segment rinses needle and tubing; the second fills a factory-calibrated shear-valve loop of 25 ± 2 µl. Sampling stops when blood reaches the optical detector (threshold about 2 M RBC). Whole-blood volume taken is roughly 150 µl. With the sampler, a second 25 ± 2 µl loop is used. First-dilution diluent is 1800 µl, so the nominal diluent-to-sample ratio is 1800 / 25 = 72 (range about 66.67 to 78.26 depending on actual loop volume stored in machine parameters).

IMAGES PREVIEW

Biocode Hycel Diana 5 Evolution Xenia Technical Manual 576CA70E April 2006 — PDF page 1 preview
Biocode Hycel Diana 5 Evolution Xenia Technical Manual 576CA70E April 2006 — PDF page 9 preview
Biocode Hycel Diana 5 Evolution Xenia Technical Manual 576CA70E April 2006 — PDF page 22 preview

Later cycles insert an isolating bubble, push the 25 µl loop with 1.775 ml of diluent into the WBC preparation chamber (1/72), homogenize that dilution while the 2.5 ml syringe aspirates HEMALYSE into the preheater, then transfer 490 µl of the first dilution into the lower 25 µl dilution loop. Concurrently the three reservoirs (sheath 1, sheath 2, diluent) are topped to their detection electrodes and the vacuum trash is drained. A second bubble sampled through valve 38 frames the RBC dilution on both sides of the valve. The dilution loop is then pushed with about 2000 µl of diluent into the RBC chamber under agitation while the WBC chamber is flushed to vacuum trash. Every later cycle in the chapter continues this style of annotated schematic: active valves darkened, fluid paths bolded, arrows for direction, so a technician can watch a live sequence and name the failing element.

Measurement Science, Artefacts, and Scattergrams

The technical-discussion chapter explains why hydrodynamic focusing is used instead of a traditional aperture-only hematology path, then walks through internal quality control, moving averages, calibration and cleaning, and reference-value handling. System-error notes cover missed blood detection, pressure faults, and RBC/WBC counting kinetics. Biological-artefact pages list the usual high/low traps for each reported parameter: cryoglobulin and giant platelets inflating RBC; cold agglutinins and in-vitro hemolysis dropping RBC or Hb; carboxyhemoglobin above 10 %, lipemia, hyperbilirubinemia, and WBC above 30 giga/L raising Hb; hyponatremia raising Hct; hyperglycemia and reduced RBC deformability raising MCV; MCH near the ISLH figure of 30.5 pg; MCHC near 33–34 % with values above 36 % treated as artefact or MCV bias (saturation given as 37 %). Software version V5.1 flags Hb, MCH, and MCHC as suspect when WBC exceeds 50 giga/L.

WBC work covers physical measures, chemical preparation, cells counted, and the differentiation algorithm (eosinophils, abnormal scattergrams, NRBC/platelet aggregates, debris, immature cells, density scattergram, basophils, resistant RBC). Limits of differentiation produce NIL (Non Identified Leucocytes) when basophils or monocytes cannot be estimated; mono- and bidirectional host transfer rules for NIL are spelled out, including Hemacell Plus compatible mode. Morphological alarms raise a microscope icon and mark the differential Positive: debris, Plasmodium, NRBC, resistant RBC/erythroblasts, and related flags. Quantitative alarms, default versus positive validation, and result transmission close the chapter. A long atlas then shows five-part differential scattergrams, suspicious alerts, LMN histograms, calibrator and control plots, RBC, artefact, platelet, WBC, and eosinophil pictures, plus handling-error and abnormal-differentiation examples a bench tech can match to printed reports.

Fluidic Hardware and Maintenance Software

The fluidics volume is an illustrated parts map rather than a narrative. Analyzer fluidics include high pressure and high vacuum, decanter cleaning, internal and external sheath regulators, stand-by versus in-cycle pressure and vacuum, sheath pressures, the SERCOM pump and atmospheric trash, internal and external sheath reservoirs with a written sheath-reservoir cleaning procedure, the diluent reservoir, and the sheath-1, external-sheath, and diluent filters. Compressor-housing fluidics, new-compressor connections, and rear-panel external connections follow. Name-sorted and reference-sorted indexes plus a complete fluidic diagram let a tech find a Lee valve, Cavro three-way, minisol, or filter by either wording or Li number.

Maintenance software is organized by module. After log-in, the fluidic module offers Tests, Embase, Prime/Drain, Research of contaminations, and a Miscellaneous window. The power module exposes counting status, ADC/DAC, TTL I/O, stirring motors, and detectors. Software adjustments edit the parameters file, gains, and thresholds. Sampling and sampler pages drive motors and detectors. Acquisition-module pages teach histogram and scatter setup, then how to read kinetic tables, volume histograms, and color scattergrams inside the service program. Cytometer work is a full mechanical procedure: cover and injector removal, injector and ruby-holder cleaning, reinstallation, and complete cytometer exchange. After exchange, 5 µm latex particles pre-align DC and laser channels; final WALS (wide-angle light scatter) alignment uses control blood so the neutrophil cluster Y coordinate matches the control sheet (typical WALS 30 to 32). Preamplifier pots P1 (DC) and P2 (laser) are adjusted so the scatter spot sits at Y = 115 ± 1. Matching-cell percentage must exceed 75 % of the DC count and Y standard deviation should be 10 ± 3; out-of-range figures point to sheath-pressure error, dirty aperture, or a cytometer that must be swapped.

Alarms, Scheduled Service, and Electronics

Troubleshooting is grouped the way the instrument reports trouble: reagent alarms (Hematon 5, Hemacore, Hemalyse, trash), fluidic alarms (pressure, vacuum), measure alarms (WBC, Hb, abnormal gain on RBC/WBC/Plt, temperatures, blood detectors), mechanical alarms (BSV, turret, piercing needle, sampling needle, dilutors), and communication/LIS alarms. Practical checks include backup-battery voltage above 3 V (3.6 V listed), +5 V / ±15 V rails with +5 V raised to 5.2 V if needed, RAM-clear by unseating boards, and setting the serial protocol to NONE when no LIS is attached so false COM-2 alarms stop.

Complete PDF Bookmark Tree

Complete Bookmarks

The complete bookmarks in the “Biocode Hycel Diana 5 Evolution Xenia Technical Manual 576CA70E April 2006” are as follows:

1 Installationp. 9
1.1 Environmentp. 9
1.1.1 Power supplyp. 9
1.1.2 Temperature and humidityp. 10
1.1.3 Dimensions and weightp. 10
1.2 Physical installationp. 11
1.2.1 Unpackingp. 11
1.2.2 Diana Évo / Xénia reagents installationp. 12
1.2.3 Compressorp. 13
1.3 Required configurationp. 14
1.3.1 Minimum configurationp. 14
1.3.2 Recommended configurationp. 14
1.4 Installation procedurep. 15
1.4.1 Windows NT installationp. 15
1.4.2 Diana Évo / Xénia software installationp. 15
1.5 «Compile» programp. 18
1.6 First startupp. 19
1.6.1 Keywords:p. 19
1.6.2 Program preferencesp. 20
2 Fluidic cyclesp. 21
2.1 Counting cyclesp. 21
2.2 Count sequence (details)p. 56
3 Diana Évo / Xénia technical discussionp. 59
3.1 Flow cytometry principlesp. 62
3.2 Fluidics and hydrodynamic focusingp. 62
3.3 Limitations on traditionnal hematology systemsp. 63
3.4 Hydrodynamic focusingp. 64
3.5 Internal quality controlp. 69
3.6 Controlsp. 69
3.7 Moving averagep. 70
3.8 Calibration and cleaning.p. 73
3.9 Other parameters (reference values).p. 73
3.10 System errorsp. 74
3.11 The blood is not detectedp. 74
3.12 Pressuresp. 74
3.13 RBC and WBC counting kineticsp. 75
3.14 Biological artefactsp. 76
3.15 White Blood Cells (WBC)p. 76
3.16 Red Blood Cells (RBC)p. 77
3.17 Haemoglobin (Hb)p. 77
3.18 Hematocrit (Hct)p. 77
3.19 Mean Cellular Volume (MCV)p. 77
3.20 Mean Corpuscular Haemoglobin (MCH)p. 77
3.21 Mean Corpuscular Haemoglobin Concentration(MCHC)p. 77
3.22 Platelets (Plt)p. 78
3.23 WBC differentiationp. 79
3.24 Physical Measuresp. 79
3.25 Chemical preparationp. 79
3.26 Number of WBC analysedp. 79
3.27 Differentiation algorithm principlep. 80
3.28 Limits of the differentiation (NIL)p. 82
3.29 Morphological alarmsp. 82
3.30 Quantitative alarmsp. 83
3.31 Validate / reject a resultp. 84
3.32 Validation by defaultp. 84
3.33 Positive validationp. 84
3.34 Transmission of resultsp. 84
3.35 Glossaryp. 84
3.36 White blood cells scattergramsp. 85
3.37 Five part differential scattergramp. 85
3.38 Suspicious alertsp. 85
3.39 LMN histogramp. 87
3.40 Scattergramsp. 89
3.41 Calibratorsp. 89
3.42 Controlsp. 90
3.43 Red blood cellsp. 95
3.44 Artefactsp. 100
3.45 Plateletsp. 104
3.46 White blood cellsp. 108
3.47 Eosinophilsp. 110
3.48 Handling errorsp. 116
3.49 WBC abnormal differenciationp. 119
4 Fluidicsp. 139
4.1 Componentsp. 141
4.2 Analyzer fluidicsp. 144
Fluidic1 : High pressurep. 144
Fluidic2 : High vacuump. 146
Fluidic3 : Internal sheath regulatorp. 148
Fluidic4 : External sheath regulatorp. 150
4.3 1 High pressure and vacuum regulationsp. 152
4.4 2 Pressure and vacuum in stand-byp. 153
4.5 3 Pressure and vacuum during cyclesp. 154
4.6 4 Sheath pressuresp. 155
Fluidic5 : SERCOM pump / atmospheric trashp. 156
Fluidic6 : Reservoir internal sheathp. 158
Fluidic7 : Reservoir external sheathp. 160
Fluidic8 : Reservoir diluentp. 162
Fluidic9 : Filter sheath 1p. 164
Fluidic10 : External sheath filterp. 166
Fluidic11 : Diluent filterp. 168
Fluidic12 : Vacuum trashp. 170
Fluidic13 : Atmospheric trashp. 172
Fluidic14 : Pressure Hematon 5p. 174
Fluidic15 : Dilutors outputsp. 176
Fluidic16 : Piercing needle, sampling needlep. 178
Fluidic17 : Washing wellp. 182
Fluidic18 : V10 – V12 / BSVp. 184
Fluidic19 : V16 – V17 / BSVp. 186
Fluidic20 : V11 / BSVp. 188
Fluidic21 : Preheaterp. 190
Fluidic22 : RBC / WBC transfer to cytometerp. 192
Fluidic23 : Hemalyse 5 sub-basep. 196
Fluidic24 : Preparation chambers, sampling valvep. 198
Fluidic25 : Cytometerp. 204
4.7 Compressor housing fluidicsp. 206
Fluidic26 : Connections on compressorp. 206
Fluidic27 : Connection on vacuum reservoirp. 207
Fluidic28 : Connections on pressure reservoirp. 208
4.8 Connections on new compressorp. 209
Fluidic29 : Connections on compressorp. 209
Fluidic30 : Connection on vacuum reservoirp. 210
Fluidic31 : Connections on pressure reservoirp. 211
4.9 External connectionsp. 212
Fluidic32 : External trash fluidics + External check valves fluidics (Li 21711846)p. 212
Fluidic33 : HEMACORE fluidics(Li 21711847)p. 213
Fluidic34 : HEMALYSE 5 fluidics (Li 21711849)p. 214
Fluidic35 : HEMATON 5 fluidics (Li 21711847)p. 215
4.10 Index sorted by namep. 216
4.11 Index sorted by referencep. 219
4.12 Fluidic diagramp. 222
5 Maintenance softwarep. 223
5.1 Acces and log-inp. 224
5.2 Necessary tools and partsp. 225
5.3 Software organisationp. 225
5.4 Fluidic modulep. 226
5.4.1 Maintenance windowp. 226
5.4.2 «Tests»p. 227
5.4.3 «Embase»p. 228
5.4.4 «Prime / Drain»p. 235
5.4.5 «Research of contaminations»p. 235
5.4.6 «Window Miscellaneous»p. 242
5.5 Power modulep. 242
5.5.1 Counting statusp. 242
5.5.2 ADC / DACp. 243
5.5.3 TTL input / outputp. 244
5.5.4 Stirring motorsp. 245
5.5.5 Detectorsp. 245
5.5.6 Miscellaneousp. 246
5.6 Software adjustmentsp. 247
5.6.1 Parameters filep. 247
5.6.2 Gain / Thresholdsp. 248
5.6.3 Miscellaneousp. 248
5.7 Sampling modulep. 249
5.7.1 Motorsp. 249
5.7.2 Miscellaneousp. 250
5.8 Samplerp. 251
5.8.1 Motors / detectorsp. 251
5.8.2 Miscellaneousp. 251
5.9 Acquisition module adjustmentsp. 252
5.9.1 Introduction and theoryp. 252
5.9.2 Adjustmentsp. 254
5.10 Acquisition histogramp. 257
5.10.1 Acquisition scatterp. 259
5.10.2 End of adjustmentp. 261
5.11 Interpretation of datas in maintenance programp. 262
6 Cytometerp. 267
6.1 Cytometer cleaning (routine maintenance)p. 268
6.2 Dismountingp. 268
6.3 Cover disassemblyp. 269
6.4 Injector disassemblyp. 270
6.5 Injector cleaningp. 271
6.6 Ruby holder dismountingp. 272
6.7 Ruby holder cleaningp. 272
6.8 Reinstallationp. 273
6.9 Replacementp. 273
6.10 Parts, necessary tools:p. 273
6.11 Cytometer exchangep. 274
6.12 Acquisition module adjustmentsp. 275
6.13 Introduction and theoryp. 275
6.14 Adjustmentsp. 277
6.15 Acquisition histogramp. 280
6.16 Acquisition scatterp. 282
6.17 End of adjustmentp. 284
7 Troubleshooting on Diana Évo / Xéniap. 285
7.1 Preliminaryp. 286
7.2 Reagent alarmsp. 286
7.2.1 Hematon 5p. 286
7.2.2 Hemacorep. 287
7.2.3 Hemalysep. 287
7.2.4 Trashesp. 287
7.3 Fluidic alarmsp. 288
7.3.1 Pressuresp. 288
7.3.2 Vacuump. 288
7.4 Measures alarmsp. 289
7.4.1 Alarms on WBCp. 289
7.4.2 Alarms on RBCp. 290
7.4.3 Alarms Hbp. 290
7.4.4 Abnormal gain on RBC, WBC and Pltp. 292
7.4.5 Alarms temperaturesp. 292
7.4.6 Alarms blood detectorsp. 292
7.5 Mechanical alarmsp. 293
7.5.1 BSV motorp. 293
7.5.2 Turret motorp. 293
7.5.3 Piercing needle motorp. 293
7.5.4 Sampling needle motorp. 293
7.5.5 Dilutorsp. 294
7.6 Communication alarmsp. 294
7.6.1 Sequen ces transmissionp. 294
8 Maintenance schedulesp. 297
8.1 Necessary partsp. 297
8.2 Semi-annual maintenance schedulep. 299
8.3 Yearly maintenance schedulep. 300
8.4 Biennial maintenance schedulep. 301
9 Wiringp. 303
9.1 Sheath connection board to mother boardp. 303
9.2 Sampling connection board to mother boardp. 304
9.3 Reset wiringp. 305
9.4 Cuves connection board to mother boardp. 306
9.5 Cytometer connection board to mother boardp. 307
10 Mother board schemasp. 309
11 Sampling board schemasp. 317
12 Power P1 (logic) schemasp. 335
Page 1p. 338
Page 2p. 340
Page 3p. 342
Page 4p. 344
Page 5p. 345
Page 6p. 346
13 Sequence board schemasp. 349
Page 1p. 352
Page 2p. 354
Page 3p. 356
Page 4p. 358
Page 5p. 360
Page 6p. 362
Page 7p. 364
14 Power P2(analogic) schemasp. 367
15 Counting board schemasp. 377
16 Preamplifier schemasp. 401
17 Hemoglobine preamplifierp. 409
17.1 Exchangep. 409
17.1.1 Adjustment of hemoglobin preamplifier ZEROp. 409
17.1.2 Automatic adjustment of hemoglobin gainp. 410
17.2 Schematicsp. 411
18 Power supplyp. 413
18.1 Diagramp. 413
18.2 Implantationp. 414

Three printed schedules—semi-annual, yearly, and biennial—are built around bleach cleaning of the shear valve and loops, cytometer, needles, rinsing well, decanters, vacuum trash, valves S6/S25/S26/S29/S21/S22, valve sub-bases, and WBC/RBC transfer paths, plus pressure/vacuum verification, platelet background, impedance gain on 5 µm latex, laser WALS printouts, and a seven-replicate precision run on fresh normal blood. Biennial extra work replaces pressure/vacuum filters, dirty injection tubes, waste check valves, syringe tips, selected minisol and microsol valves, HEMATON and HEMACORE check valves, piercing and sampling needles, pressure regulators, Lee valves S40 and S41 on the WBC line, Cavro three-way valves, and the 1 ml and 2.5 ml dilutor syringes. Power-supply targets listed on that schedule are +5.2 V, ±15 V, +28 V, and +24 V. A two-year parts table gives Li references for silencers, check valves, filters, needles, regulators, FAS seals, and related hardware.

Closing chapters print wiring between sheath, sampling, reset, cuves, and cytometer boards and the mother board, then full schematics for the mother board, sampling board, Power P1 logic, sequence board, Power P2 analog, counting board, preamplifier, and hemoglobin preamplifier (zero and automatic gain procedures). A power-supply diagram and implantation drawing finish the book.

Chapters Covered

  • Installation — Environment (power supply, temperature and humidity, dimensions and weight); Physical installation (unpacking, reagents, compressor); Required PC configuration; Installation procedure and Compile program; First startup and program preferences
  • Fluidic cycles — Counting cycles from sampling through dilution, bubble isolation, RBC/WBC chamber work, transfer to the cytometer, and cleaning
  • Diana Évo / Xénia technical discussion — Flow cytometry and hydrodynamic focusing; Internal quality control, moving average, calibration and cleaning; System errors; Biological artefacts on WBC, RBC, Hb, Hct, MCV, MCH, MCHC, platelets; WBC differentiation algorithm, NIL limits, morphological and quantitative alarms; Validation and result transmission; Scattergram atlas (five-part differential, alerts, calibrators, controls, artefacts, handling errors)
  • Fluidics — Components; Analyzer fluidics (high pressure/vacuum, sheath regulators, stand-by and cycle pressures, reservoirs, filters); Compressor housing; External connections; Name and reference indexes; Fluidic diagram
  • Maintenance software — Access; Fluidic, power, sampling, and sampler modules; Software and acquisition-module adjustments; Interpretation of maintenance data
  • Cytometer — Routine cleaning; Dismounting injector and ruby holder; Replacement and cytometer exchange; Acquisition adjustments after exchange
  • Troubleshooting on Diana Évo / Xénia — Reagent, fluidic, measure, mechanical, and communication alarms
  • Maintenance schedules — Necessary parts; Semi-annual, yearly, and biennial checklists
  • Wiring and board schematics — Sheath, sampling, reset, cuves, and cytometer connections; Mother, sampling, Power P1, sequence, Power P2, counting, and preamplifier drawings
  • Hemoglobin preamplifier — Exchange, zero and automatic gain, schematics
  • Power supply — Diagram and implantation

Taken together, the Biocode Hycel Diana 5 Evolution Xenia Technical Manual 576CA70E April 2006 gives a field engineer the sequence files, fluid maps, optical alignment numbers, alarm trees, and board drawings needed to put the analyzer on the bench and keep counts and differentials inside specification.

Use This Manual on the Bench

With this file a technician can site the analyzer and compressor, plumb HEMATON-5, HEMACORE, HEMALYSE-5 and waste, compile instrument sequences, watch each counting cycle against the printed fluid diagrams, interpret five-part scattergrams and NIL/morphology flags, clean or exchange the cytometer and shear-valve loops, set DC and laser gains against latex and control blood, work reagent and pressure alarms in the order the software raises them, and run the semi-annual through biennial part list. Following the printed volumes, ratios, WALS targets, and rail voltages matters because a dirty sheath regulator, a mistimed bubble, or a drifted P2 pot changes kinetics and can shift MCV, platelet background, or the neutrophil cluster before any numeric flag appears.

Delivery is an instant PDF download after payment. There is no shipped binder. Open the file on a workshop PC or tablet and print the cycle drawings, maintenance checklists, or board schematics you need for the job.

Reviews

There are no reviews yet.

Be the first to review “Biocode Hycel Diana 5 Evolution Xenia Technical Manual 576CA70E April 2006”

What Our Customers Say

★★★★★ Live reviews from customers
Loading customer reviews...
0
    0
    Your Cart
    Your cart is emptyReturn to Shop
    🛒
    Recently Purchased
    🕒 verified order